    SB-800 HQ Polymer-based packed columns for aqueous SEC (GFC) analysis Supports a wide range of molecular weight sample analysis The eluent can be replaced with DMF (except SB-802 HQ and SB-807 HQ), enabling the analysis of polar polymers Method using […]

    shodex-image

    Please refer to the following tables for the classification of solvents in terms of capability of being replaced with original in-column solvent. However, frequent replacement of in-column solvent may shorten column life, and is, therefore, not recommended. Refer to Solvent Exchange […]

    shodex-image

    It is well known that ion analysis is used with conductivity detector. There are two methods for conductivity detectors: the suppressed conductivity method and the non-suppressed conductivity method. The Non-suppressed conductivity method can be easily performed by adding a conductivity […]

    shodex-image

    Two ion chromatography methods are well known: the suppressed method and the non-suppressed method. The suppressed method is suitable for anion analysis. Sodium carbonate aqueous solution is used as the eluent of Shodex columns for the suppressed method. The suppressor […]

    shodex-image

    Generally, anion and cation analysis are used under ion exchange mode with a conductivity detector. There are two methods for conductivity detectors: the suppressed conductivity method (suppressor method) and the non-suppressor method (non-suppressed method). In addition to columns for ion […]

    shodex-image

      There are two types of ion chromatography: non-suppressor method and suppressor method. While the suppressor method requires a suppressing module, the non-suppressor method can be used for ion analysis by simply connecting an electro-conductivity detector to an all-purpose LC […]

    shodex-image

      Two methods are well known as ion chromatography measurement : the suppressed conductivity method and the non-suppressed conductivity method. In the suppressed conductivity method, a special device (suppressor) installed after the analytical column is needed for removing the opposite […]