    To exchange/replace the solvent currently filled in the Shodex GPC series columns, follow the below solvent exchange procedures. Note: When replacing the solvent, set the column temperature at 40 to 60 ℃ and start the system at less than half […]

    shodex-image

    Please refer to the following tables for the classification of solvents in terms of capability of being replaced with original in-column solvent. However, frequent replacement of in-column solvent may shorten column life, and is, therefore, not recommended. Refer to Solvent Exchange […]

    shodex-image

    Polystyrene standards were analyzed using GPC HK-402, an organic SEC (GPC) column for ultra-rapid analysis. The RI chromatograms showed that n-propylbenzene and benzene peaks overlapped with eluent-related negative peaks. RI Sample : 0.5 % each, 5 µL 1. Polystyrene (Mp: […]

    shodex-image

    Polystyrene standards were analyzed using GPC HK-403, an organic SEC (GPC) column for ultra-rapid analysis. The RI chromatograms showed that n-propylbenzene and benzene peaks overlapped with eluent-related negative peaks. RI Sample : 0.5 % each, 5 µL 1. Polystyrene (Mp: […]

    shodex-image

    Polystyrene standards were analyzed using GPC HK-406, an organic SEC (GPC) column for ultra-rapid analysis. Comparing the separation between an HK-406 with a conventional SEC column of equivalent at the flow rate of 1.0 mL/min, HK-406 achieved the same level […]

    shodex-image

    Polystyrene standards (Mp580) were analyzed using GPC HK-401, an organic SEC (GPC) column for ultra-rapid analysis. Comparing the separation between an HK-401 with a conventional SEC column of equivalent at the flow rate of 1.0 mL/min, HK-401 achieved the same […]

    shodex-image

    For measurement of molecular weight distribution by SEC mode, it is necessary that no interaction exists between packing material and the sample. Therefore it is important the eluent in which any interaction does not exist between packing material and sample, […]

    shodex-image

      GPC (Gel Permeation Chromatography) separates solutes in a sample according to molecular size of solutes. Pores of packing materials are used for the separation. When performing GPC, it is assumed that no interaction between solutes and packing material exists. […]

    shodex-image

    The solubility of solvents which are commonly used for GPC analysis is shown below:

    shodex-image

    Physical chemistry properties of solvents which are commonly used for GPC analysis is shown below. Solvent Chemical formula Boiling point (°C) Viscosity cp, (25°C) Transmission limit (nm) Refractive index nD25 Acetic acid CH3COOH 118 1.1 – 1.372 Acetonitrile CH3CN 82 […]