Calibration curves for the GPC HK-400 series, and organic SEC (GPC) columns for ultra-rapid analysis, using PMMA are shown. Sample : Poly(methyl methacrylate) Columns : Shodex GPC HK-400 series (4.6 mm I.D. x 150 mm each) Eluent : 5 mM […]

Polystyrene (Mp 580) was analyzed using GPC HK-406, an organic SEC (GPC) column for ultra-rapid analysis. Different numbers of HK-401 columns (1, 2, and 3 units) were used alone/together in series to compare their separation efficiencies. The use of multiple […]

Polystyrene standards were analyzed using GPC HK-402, an organic SEC (GPC) column for ultra-rapid analysis. The RI chromatograms showed that n-propylbenzene and benzene peaks overlapped with eluent-related negative peaks. RI Sample : 0.5 % each, 5 µL 1. Polystyrene (Mp: […]

Polystyrene standards were analyzed using GPC HK-403, an organic SEC (GPC) column for ultra-rapid analysis. The RI chromatograms showed that n-propylbenzene and benzene peaks overlapped with eluent-related negative peaks. RI Sample : 0.5 % each, 5 µL 1. Polystyrene (Mp: […]

Polystyrene oligomer was analyzed using organic solvent SEC (GPC) semi-micro columns, 2x KF-402HQ. Poly-styrene standards separated had average MW of 580, contained dimer (MW:266) to octamer (MW1098). Sample : Styrene oligomer Column : Shodex GPC KF-402HQ (4.6mmID*250mm) x 2 Eluent […]

Figure 1 shows the relationships between the flow rate and the theoretical plate number (TPN) when using an organic solvent SEC (GPC) semi-micro column, KF-402HQ. With flow rate of over 0.2 mL/min, TPN of over 30,000 was obtained. While with […]

The effect of flow rate was examined for organic solvent SEC (GPC) semi-micro columns, KF-402HQ and KF-403HQ. Improved separation is observed at lower flow rate, especially for the analysis of large MW compound such as EPON1001. Please see Effects of […]

Figure presents the relationship between the concentration of sample and the height equivalent theoretical plate (HETP) for organic solvent SEC (GPC) semi-micro columns, KF-402HQ Lower HETP is obtained with sample concentration Effects of sample concentration (2) (KF-403HQ + KF-402HQ) for […]

The effects of sample concentration on separation of EPON1001 polymer were examined using organic solvent SEC (GPC) semi-micro columns, KF-403HQ and KF-402HQ No considerable decline in resolution was observed up to analysis of 5% (m/v) EPON1001. Please refer to Effects […]

The effect of column temperature on HETP was examined for organic solvent SEC (GPC) semi-micro column, KF-402HQThere is not a considerable temperature influence for the analysis of relatively small MW compound such as dicyclohexylphthalate. However, improved separation can be expected […]