Hydrogenated maltose starch syrup was analyzed on SUGAR KS-801 according to Japanese Pharmaceutical Excipients 2013 method. It is necessary for the choice of column to satisfy resolution of maltitol/propylene glycol (internal standard) of ≥ 1.5. It was confirmed that resolution was more than 10.0 when they were analyzed using KS-801.
Sample : 1% each, 10µL
1. Maltitol
2. Propylene glycol
Column : Shodex SUGAR KS-801 (8.0mmI.D. x 300mm) Eluent : H2O Flow rate : 0.37mL/min Detector : Shodex RI Column temp. : 50℃