According to the US Pharmacopeia (USP), analysis of maltose should be done using a column filled with L58 packing materials. The system suitability requires to satisfy the resolution between maltose and maltotriose to be ≥ 1.6 and relative standard deviation (RSD) of the maltose peak area to be ≤ 2.0%.
SUGAR KS-801 was confirmed to satisfy all required conditions for the analysis.
Sample : 10 mg/mL each
(in H2O), 20 µL
1. Maltotriose
2. Maltose
3. Glucose
Column : Shodex SUGAR KS-801 (8.0 mm I.D. x 300 mm) Eluent : H2O Flow rate : 0.35 mL/min Detector : Shodex RI Column temp. : 80 °C