IC YS-50 is a column ideal for simultaneous analysis of monovalent and divalent cations. Trivalent chromium and aluminium ions can be also analyzed by increasing the concentration of methanesulfonic acid in the eluent.
Sample : 20µg/mL each, 20µL
1. K+
2. Ca2+
3. Cr3+
4. Al3+
Column : Shodex IC YS-50(4.6mmI.D. x 125mm) Eluent : 12mM Methanesulfonic acid aq. Flow rate : 1.0mL/min Detector : Non-suppressed conductivity Column temp. : 40°C