Sample :
Cr: Creatinine
U: Uric acid
Columns : C1 : Shodex Asahipak GS-320 7C (7.6mmID*100mm)
C2 : Shodex Asahipak ES-502C 7C (7.6mmID*100mm)
C3 : Dowex 50-X8
C4 : Shodex Asahipak ES-502N 7C (7.6mmID*100mm)
C5 : Shodex Asahipak GS-310 7G (7.6mmID*500mm)
Eluent : 60mM Propionic acid + 15mM Succinic acid + 60mM NaOH in 1kg H2O (for C1, C2, C4, C5)
50mM Propionic acid + 15mM Succinic acid + 60mM NaOH in 1kg H2O (for C3)
Flow rate : 1.2mL/min
Detector : UV(234nm,290nm)
Column temp. : 30°C (for C1),40°C (for C2, C4, C5),45°C (for C3)
Courtesy of Dr. Seki, Osaka University