Lithium tetrafluoroborate, lithium hexafluorophosphate and lithium perchlorate are used as an electrolyte of a lithium ion secondary battery. In this application, the anions of these salts were analyzed using IC SI-90 4E, an anion analysis column for non-suppressor method.
Sample : 10μL
1. F– 2mg/L
2. BF4– 5mg/L
3. ClO4– 10mg/L
4. PF6– 30mg/L
Column : Shodex IC SI-90 4E (4.0mmI.D. x 250mm) Eluent : 4.0mM Na2CO3 + 1.7mM NaHCO3 aq. Flow rate : 1.5mL/min Detector : Suppressed coductivity Column temp. : 45°C