PROTEIN LW-403 4D is a rapid semi-micro column for aqueous SEC (GFC) suitable for antibody protein analysis. A comparison of IgG aggregate separation with a conventional device and a semi-micro type device was completed. The resolution of the conventional device is reduced due to the diffusion of the sample outside the column. We recommend using a semi-micro type device to achieve the best column performance.
Sample : 5 μL
IgG from human serum 2 mg/mL
1. Aggregates
2. Trimer
3. Dimer
4. Monomer
Semi-micro type | Conventional type | ||
---|---|---|---|
Tubing | 0.1 mm I.D. | 0.25 mm I.D. | |
 UV detector | Cell volume | 2.5 μL | 18 μL |
Optical path length | 5 mm | 10 mm |
Column : Shodex PROTEIN LW-403 4D (4.6 mm I.D. x 150 mm) Eluent : 50 mM Sodium phosphate buffer (pH7.0) + 0.3 M NaCl Flow rate : 0.35 mL/min Detector : UV (280 nm) Column temp. : Room temp.