Since Asahipak NH2P-50 column is chemically stable,
the durability against acidic solvents is better than that of silica-based amino columns.
Testing procedure:
(1) Testing (63hrs)
|
(2) Replacement (1hr)
|
(3) Equilibrium (1hr)
|
Comparison between before and after testing:
Sample | Before testing | After testing | ||
---|---|---|---|---|
Elution volume (mL) |
Plate number | Elution volume (mL) |
Plate number | |
1.Fructose | 6.73 | 9600 | 6.70 | 8400 |
2.Glucose | 8.59 | 7900 | 8.58 | 8500 |
3.Sucrose | 11.88 | 9600 | 11.93 | 9300 |
4.Maltose | 14.24 | 8500 | 13.90 | 8600 |
Column : Shodex Asahipak NH2P-50 4E (4.6mmID*250mm)
Eluent : CH3CN/H2O=75/25
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 30°C