Since the packing material of Asahipak NH2P-50
column is polymer, it is durable against alkaline solvents. To the contrary,
silica-based amino columns are not durable against alkaline solvents.
Testing procedure:
(1)Equilibrium(1hr) Conditions before testing |
(2) Analysis |
(3)Testing(160hrs) |
(4)Replacement(0.5hr) |
(5)Replacement(0.5hr) |
(6) Analysis |
Comparison between before and after testing:
Sample | Before testing | After testing | ||
---|---|---|---|---|
Elution volume (mL) |
Plate number | Elution volume (mL) |
Plate number | |
1.Fructose | 7.52 | 7100 | 7.62 | 8100 |
2.Glucose | 9.63 | 6100 | 9.68 | 7700 |
3.Sucrose | 13.41 | 8400 | 13.40 | 9700 |
4.Maltose | 16.16 | 7200 | 16.08 | 7400 |
Column : Shodex Asahipak NH2P-50 4E (4.6mmID*250mm)
Eluent : CH3CN/H2O=75/25
Flow rate : 1.0mL/min
Detector : Shodex RI
Column temp. : 30°C