Polystyrene standards were analyzed using GPC HK-406, an organic SEC (GPC) column for ultra-rapid analysis. Comparing the separation between an HK-406 with a conventional SEC column of equivalent at the flow rate of 1.0 mL/min, HK-406 achieved the same level of separation with 1/6 analysis time. However, when higher separation is desired, the use of a lower flow rate is recommended.
Sample : 0.5 %, 5 µL
1. Polystyrene (Mp: 609,000)
2. Polystyrene (Mp: 53,000)
3. Polystyrene (Mp: 3,180)
Columns : Shodex GPC HK-406 (4.6 mm I.D. x 150 mm) x 2 Eluent : THF Flow rate : 1.0 mL/min, 0.6 mL/min, 0.3 mL/min Detector : RI (small cell volume) Column temp. : 30 °C