Calibration curves for Asahipak GF-HQ series using poly(sodium methacrylate) are shown. Since Asahipak GF-HQ series can be used with both aqueous and organic solvents as the eluent, it is suitable for the analysis of hydrophilic polymers such as sodium poly(styrene sulfonate) or poly(sodium methacrylate). Several types of hydrophilic polymer standard are commercially available. It is recommended to use standard polymer similar to the analyte for the calibration curve.
Sample :
Poly(sodium methacrylate) standard
Columns : Shodex Asahipak GF-HQ series (7.5mmI.D. x 300mm each) Eluent : 50mM Na2HPO4 aq./CH3CN=90/10 Flow rate : 0.6mL/min Detector : Shodex RI Column temp. : 40°C