Features of GPC
1) No interaction between solutes and packing material is assumed.
2) To prevent the interaction, selection of appropriate eluent is essential.
3) The order of elution is the order of molecular size of solutes (from larger to smaller molecular size.)
4) The solutes whose molecular size is larger than the pore size of packing material cannot be separated each other. The upper limit of the molecular size which can be separated is called “exclusion limit”. The exclusion limit is peculiar to column type.
5) Elution volume which corresponds to exclusion limit is called “void volume(Vo)”, which is a very important characteristic of GPC column.
6) The total volume of pores is called “inner volume(Vi)”. The elution volumes of all solutes by pure GPC mode must exist between Vo and Vo + Vi.
Causes which prevent GPC mode
When performing GPC mode analysis, the most important thing is how to realize pure GPC mode. For the analyses of polystyrene and saccharides, the optimal conditions to realize pure GPC mode are well-known. However, for some samples, it is difficult to find the optimal condition. And, sometimes, chromatograms which do not act on GPC mode appear or only poor reproducibility can be obtained.
When pure GPC mode is performed, the calibration curve exist in the range of Vo and Vo + Vi as shown by the solid line in the right figure. However, when pure GPC mode is not performed, the calibration curve shown by the broken line can be obtained.
In case (A), some solute elute before Vo. The possible causes of this too fast elution are as follows:
* When the sample is ionic and the eluent is not an appropriate one, this phenomenon may happen. Please refer to GPC Analysis of Ionic Sample. Even when the elution volumes are in the range of Vo and Vo + Vi, this may be the cause of poor reproducibility.
* When the aggregation of sample is taken place, this phenomenon may happen.
* When the ion-exclusion with packing material is taken place, this phenomenon may happen.
In case (B) and (C), solute must be adsorbed by the packing material. Such adsorption is mainly caused by hydrophobic or hydrophilic interaction and sometimes by hydrogen bond or ionic bond. The way t prevent such adsorption is as follows:
* To prevent hydrophobic or hydrophilic interaction, an eluent which has higher solubility to the solute and/or higher affinity to the packing material should be used.
* For some ionic polymers, addition of LiCl or LiBr increases the solubility and may solve the problem.
* When the sample is a polymer which has weak ionic group such as -COO–, the affinity of the polymer can be largely affected by the degree of dissociation. In such case, the degree of dissociation should be stabilized by using some buffer.