2010-12-01 (Wed) 02:02:33 Shodex Technical question

Record: 89
Category: Specifications Product: SB-806 Sample:
Question What precisely is the difference between the SB-806 and SB-806M media?     Answer OHpak SB-806HQ column is packed with the single gel having 1500A pore size. OHpak SB-806M HQ column is packed with some kinds of SB-800HQ gels for the wider molecular weight range. So, the linearity range of the calibration curve is different. Please refer to the following page as a reference. The linearity at the lower molecular weight range of SB-806MHQ is better than that of SB-806HQ. http://www.shodex.com/english/dc060208.html     Additional Comments