2011-06-23 (Thu) 23:27:40 Shodex Technical question

Record: 303
Category: Specifications Product: ODP-50 Sample:
Question Is ODP-50 compatible with THF?     Answer We are sorry but we do not guarantee the performance of ODP-50 with use of THF. However, we have used ODP-50 column with the following eluent condition for an analysis, and the column was not deteriorated with this condition. CH3CN/THF/H2O=45/20/35 So, we think THF is usable for ODP-50, though this condition in not guaranteed.     Additional Comments