2012-04-10 (Tue) 01:31:01 Shodex Technical question

Record: 590
Category: Specifications Product: KS Sample:
Question I am ordering a column, KS-801, and also need a guard column. I have not found a holder for the guard column on your webpage, so I am trying to determine whether the guard column holders that I have will work. I have one from Bio-Rad and one from Waters.     Answer Shodex guard columns are not the cartridge type but the packed type column. So, Shodex guard columns are not usable with the guard column holder. However, we think the fitting of the guard column and a column holder is the same, even if the column dimension is different. And the connection type of Shodex SUGAR KS-G is the same as that of KS-801. The connection type of KS-801 and KS-G is Internally-threaded type,No.10-32 UNF (stainless steel). Could you please confirm if the fitting between the pump and the column for the guard column? We will attach the picture of KS-G as a reference.     Additional Comments