Record: 409


Category: Column Selection
Product: SB
Sample:


Question
With glycerol bonded size exclusion support on 7 µ , 500Å spherical silica designed for the rapid analysis of proteins, carbohydrates, nucleic acids, and other water-soluble polymers. This support is appropriate for globular molecules with MW ranges from 40-2000 kD and linear molecules with MW ranges from 10-350 kD.

 
 

Answer
SEC column with 500A pore size will not be able to analyze protein having 2000 KDa nor linear molecules with 350 kDa (the pore size is too small).

I would recommend OHpak SB-804 HQ or SB404-4E.
The base material is polyhydroxymethacrylate and both having pore size of 2,000A providing an analysis range between 10,000-1000,000Da (pullulan) If higher resolution is required for smaller range compound, it may require to combine SB-804 HQ with SB-802.5 HQ. It is depending on the customer’s target analyte.
SB-804 HQ F6429103
10 um, 8.0×300 mm

SB404-4E F6429114
7 um, 4.6×250 mm
SB404-4E is shorter, thus provides shorter analysis time and owing to its smaller particle size, provides higher resolution. However, the customer has to have semi-micro HPLC system, that can control the flow rate of ~0.2 mL/min.

 
 