Skip to main content
Print

An Answer

Record: 415


Category: Column Selection
Product: SB
Sample: Carboxymethylcellulose


Question
I would appreciate to receive some technical note about SB404-4E . The customer would like to know if he could analize Carboxymethylcellulose with this column. Sorry, I realized that you mentioned about MW.. so yes for that MW SB404 can be used. However, may require to add SB-402.5 if the analyte has wide MW distribution. Also please make sure that the customer has semi-micro system to be able to fully benefit SB404.

 
 

Answer
Do you know the MW of your customers sample?
For Carboxymethylcellulose, it might require larger pore size.
http://shodex.net/index.php?seitenid=1&applic=132
Unfortunately, we do not have the one with 15,000A for SB400 series.
Also to use SB400 series, the customer has to have semi-micro HPLC system that can control the flow rate of ~0.2 mL/min.

 
 
Additional Comments

image_pdfDownload or Print PDF
Shopping Cart
➤
Scroll to Top
Select more than one item for comparison.