Analysis of PE/PEO

Record: 325


Category: Column Selection
Product: SB?
Sample:


Question

 
 

Answer
I would recommend you to consider using a combination of SB-807 HQ and SB-804 HQ.

We do not have an exact application showing the separation of 4,000, 300,000 and 6,000,000 PEOs.
Below example is the one used SB-804 HQ, gives you some idea of how much separation level it can provide between 4,000 and 300,000 PEOs.
http://shodex.net/index.php?seitenid=1&applic=898

By adding SB-807 HQ (our largest pore-sized column) will help measuring 6,000,000 PEO. As you can see that the calibration curve for SB-806 HQ (sorry we do not have calibration curve for PEO measured by SB-807 HQ), the slope is very sharp for the higher MW range. Thus, there could be a bigger uncertainty for determined MW for 6,000,000 PEO. Though, there should not be a problem separating three PEOs.
http://shodex.net/index.php?seitenid=1&reiter=r_results&applic=1234

 
 
Additional Comments