Question
Customer request application Shodex column
Reverse phase column C-18 (250 mm x 10 mm, 5 mm) Analysis of microcystin x
LC-18 column (15 cm x 4.6 mm, 5 mm) Analysis of saxitoxin-STX, GTX Goniautoxinas-and C-toxins C18-4D
Allsphere column ODS-2 mM, 250 x 4.6 mm internal diameter column (Altech) Analysis cylindrospermopsins (CYL) C18-4E
HPLC column Phenomenex C18 reverse phase (Gemini 5 mm x 250 mm 4.6) Phycotoxins – Amnesia ASP C18-4E
Column of silica reverse phase (C18) of 4.6 mm x250 Phycotoxins-diarrheal C18-4E
Reverse phase column Phenomenex C8 (5 mm Luna 250 x 4.6 mm) Phycotoxins – PSP (paralytic shellfish poison) Silica 5C8 4E

 
 

Answer
1. 250mmx10mm column: is the customer looking for a C18 column for preparative scale purpose?
Are they trying to separate microcystin from something else? I feel like SEC column might be a better choice for this.

2. This is probably for the analysis of toxins causes shellfish poisoning.
That we have an application for: http://shodex.net/index.php?seitenid=1&applic=1462
The column used here was a special custom size 2B, because of the MS used as a detector. But our standard size column, DE-413 will work to separate those compounds, using the similar mobile phase mentioned in the application or this one: http://shodex.net/index.php?seitenid=1&applic=268.
If a regular C18 column is used, probably require ion-pair reagent like discussed in NIH paper below.
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC92876/
(*DE-413 is not a C18 column, polymethacrylate RP column)
Another possibility is to use ODP-50 column with alkaline conditions.
http://shodex.net/index.php?seitenid=1&applic=267

3.
I think DE-413 will be a better choice for this analysis also.
http://shodex.net/index.php?seitenid=1&applic=1475
However, is the customer already using Altech column? If they do not have any problem with the Allsphere ODS column, this is a USP L1 column and sure C18-4E can be also used.

4. Gemini is a hybrid silica column suitable for the use with a wide pH range. As you know Shodex C18 cannot be used over pH7.5, if the customer needs to use high pH condition, we would recommend ODP50 4E (pH 2-13), DE-413 (pH 2-12), or RP18-413 (pH 1-13, 4.6×150 mm). The application will be probably very similar to the one for shellfish poisoning toxins.

5. Same as 3 and 4.

6. Same as Gemini, Luna column has wider pH range. So if the customer needs to work on wide pH range, we would recommend Asahipak C8P-50 4E (pH 2-13). If the pH range is between pH 2-7.5, Silica 5C8 4E can be also used and this will provide better resolution than C8P-50 4E.

 
 