Calcium Glucoheptonate and Calcium Gluconate

Record: 329


Category: Column Selection
Product: SUGAR
Sample:


Question
They might consider this option as they really need to find a way to do this separation.
They already have some Shodex columns, and are wondering if there’s a way of doing this on one of them.
Here are the columns they have:

– Shodex NH2P 50 4D
– Shodex RSpak KC_811 6E 6um

 
 

Answer
I’m sorry to say but I don’t think neither NH2P-50 nor KC-811 will work in this case for following reasons:

The NH2P column is an amino-HILIC column. Because of the amino groups, it naturally provides alkaline condition – means this will prevent the anomers to form – not good for anomers separations.
http://shodex.net/data/downloads/Shodex_Technical_Notebook_2_NH2P50.pdf
Please see pg 5, figure 1-7.

While KC-811 is an ion exclusion mode column, and this is not so much for sugar analysis.
As I mentioned, SC1011 will have the best chance for this case.
I really hope they can try using SC1011.

 
 
Additional Comments