Skip to main content
Print

Chitopak columns

Record: 381


Category: Column Selection
Product: GFC
Sample:


Question
One of my customer is searching for an analytical column (~250 x 4.6 mm) to perform size exclusion chromatography for sugar compounds around the range of 500 – 5000 Daltons.
He found the Shodex Chitopak KQ column in our previous catalogue, but I don’t think this is still available….would you have anything else to suggest in that case?

 
 

Answer
I could recommend either KS or SB series (Direct equivalent of chitopak is SB).
If s/he is planning to use pure water as a mobile phase, KS might be the choice.
If organic solvent needs to be used to dissolve the sample or if it’s an ionic polysaccharides and required to add salt, SB might be a better choice.

KS-803 is available in 8.0×300 while SB is available in 8.0×300 mm (SB-802.5 HQ) and 4.6×250 (SB402.5 4E).
If they have a semi-micro HPLC system (that can control the flow rate of <0.3mL/min) I would recommend SB402.5-4E over KS-803 or SB-802.5 HQ. SB402.5-4E will provide the best resolution among those three. If the customer is not sure which one to choose from, would you please ask them the followings? -Type of sugar they are trying to analyze. -Is the purpose separation of several polysaccharides –if so, what will be the MW differences. Or is s/he trying to obtain MW distribution information? -Do they have semi-micro HPLC system?     Additional Comments

image_pdfDownload or Print PDF
Shopping Cart
➤
Scroll to Top
Select more than one item for comparison.
Skip to content