Chromatography column for (minimum 100000 bases long) dsDNA purification

Record: 168


Category: Column Selection
Product: SB-807
Sample:


Question
Your column looks interesting, ideally we want to separate 66 000 000 or 122 000 000 dA DNA molecules from shorter fragments – removing salts, enzymes, dNTPs, and small DNA fragments. It can be ss or dsDNA, but it’s critical that the DNA isn’t sheared much. Do you have any idea as to how large the shear forces will be in the SB-807?

 
 

Answer
We have an application using SB-807 column for the MW analysis of polyacrylamide with average MW 5,500KDa.
http://shodex.net/index.php?seitenid=1&applic=1079

Also below application shows the effects of particle sizes on shear degradation.
This is not done by SB series column, but high temperature GPC (UT) column, which is filled with 30 um-particle size gel.
The application shows that this column was feasible measuring polymer with 20MDa MW without causing shear degradation.
http://www.shodex.com/english/da060306.html
The particle size of gel used in SB-807 is 35 um, thus can be expected to have similar or less shear degradation effect compared to the above UT column example.

 
 
Additional Comments

image_pdfDownload or Print PDF