Chrome Application

Record: 338


Category: Column Selection
Product:
Sample: chromate?


Question
Do you have an application for the separation of trivalent chrome? It seems that the chrome application can be managed with the following Dionex columns: Guard Column IonPac NG1 and Analytical Column IonPac AS7. Can you match these with Shodex equivalents?

 
 

Answer
I belive this is for the analysis of chromate (CrO3 2-) in water.
Shodex equivalent for IonPac AS7 will be SI-90 4E.
Although we do not have an application showing the analysis of chromate, it will elute at later time than SO4 2- (another trivalent anion).

We also have an information on chromate analysis using our non-suppressor column I-524A.
http://shodex.net/index.php?seitenid=1&applic=948

If customer has a suppressor system, it will provide better sensitivity, so I would recommend SI-90 4E over I-524.
My one concern is that in Dionex website, they show EPA Method 218.6 (hexavalent) Chromium in waste water, using AS7 column.
http://www.dionex.com/en-us/products/columns/ic-rfic/specialty-packed/ionpac-as7/lp-73274.html
The method uses ammonium sulfate + ammonium hydroxide mobile phase, but unfortunately we cannot gurantee the performance of SI-90 4E at this condition.

 
 
Additional Comments