Question
The purpose for me to purchase these columns is to try to separate two components (50KD and 25KD). Do you think if these columns are applicable?

 
 

Answer
I believe you are analyzing protein samples?
In that case acually, using either one of KW-402.5 or KW-403 might be able to separate the two proteins.
Please see below applications showing the separation of protein standards.
Protein with 25K will fall between analyte 6 and 7, while 50K will be between 3 and 4.

KW-402.5
http://shodex.net/index.php?seitenid=1&applic=1309

KW-403
http://shodex.net/index.php?seitenid=1&applic=1310

As you can see, the separation of those two are similer in both KW-402.5 and KW-403.
In that morecular range of 10K-100K, the slope of calibration curves for KW-402.5 and KW-403 are similar, but slightly smaller for KW-403, meaning it may provide better separations.
http://shodex.net/index.php?seitenid=1&applic=1228

If you need better separation of two components, you may combine 2xKW-403. Or if your feafuture project requires you to analyze smaller MW range protein I would suggest the combinateion of KW-403+KW-402.5. But, if you are not in a hurry (have a time to wait for another column if necessary), I suggest you to try using one KW-403 to see the separation.

Few things you may need to consider when using KW-400 series is the flow rate, detector cell size, and injection volumes.
Our recommended flow rate for KW-400 series is <0.3 mL/min, thus requires a LC system provides a stable flow at this rate. Sorry I should had included this in my ealier email. Below is a detailed explanation on KW-400 series. http://shodex.net/data/downloads/Shodex_Technical_Notebook_5_KW400.pdf     Additional Comments