1.
-Tirosina is Tyrosine.
So for the separation of glutamic acid, lysine, alanine, and tyrosine, he can use CXpak P-421S with mobile phase gradient of No.3, No.4, and No.5 only.

-Trimetropin. I believe this is for the separation of trimethoprim and sulfamethoxazole.
I think RP will separate those two.
Sulfanilamides has very similar structure to sulfamethoxazole.
http://shodex.net/index.php?seitenid=1&applic=272
Since those are alkaline compounds, probably better to use polymer RP columns like DE-413 used in the application.

2. Separation of ciclidextrina B (cyclodextrine B), lutrol F127 (Poloxamer 407), enroflaxin, docyicycline (doxycycline).

Is it really for the separation of all four?
I think it’s separation of (a) cyclodextrine and poloxamer, and (b) enroflaxin and doxycycline.
(a) Probably SEC.. SB series. Need to find out the approximate MW of their targets.
(b) RP. I think regular C18 will do.

 
 
Additional Comments