Sample: Finasteride


Question
I have a question from Brazil , they want to analize , I think we can recommend our c18 column or Shodex but the ph range of this column is up to 8 , so I would suggest the ODP2 but not sure . could you please help me .
FINASTERIDE :

http://www.srcosmos.gr/srcosmos/showpub.aspx?aa=5514 HPLC ARTICLE
http://www.waters.com/waters/partDetail.htm?partNumber=WAT044375 COLUMN SPECIFICATIONS

 
 

Answer
Not because of the pH, but because of the temperature requirement -if they are going to follow the method you found, I recommend ODP40 4E.

 
 