Question
me if we have a GPC column to separate modified resins derived from rosin tree , since its used as formulation for ink and adhesives. The size is approximate 0,5 a 8000 Daltons.

 
 

Answer
For SEC analysis, it is important to know what we can use to dissolve the sample.
I think it is soluble in chloroform… if this is true we can recommend K-803L.
If the resin is soluble in THF, I would recommend KF-803L instead.

The purpose of the analysis is also important.. are they trying to “separate” different unit of the polymer? If so how much MW difference would they have?
Or is this a simple MW distribution determination?
Also need to note is that anything smaller than ~100 will come out together and won’t be able to separate using GPC.

 
 