Skip to main content
Print

Reducing Run Time for Corn Syrup Sugar Analysis with Shodex Columns

Record: 802


Category: Column Selection
Product: KS, VG
Sample: Corn Syrup


Question
A customer analyzing corn syrup currently has a 15-minute run time at 80 °C. They would like to shorten this to ~5 minutes while still resolving DP2–DP4 peaks. Could a smaller particle size Shodex column help?

Answer
Reducing to a true 5-minute run is unlikely without losing resolution of DP2–DP4. However, shorter runs (~8–10 minutes) are possible depending on method priorities:

  • SEC Option: A single KS-801 column can bring run times closer to 10 minutes, though DP3 and DP4 may not baseline-separate. Example: Sugar Alcohols (KS-801).
  • HILIC Option: VG-25 4D can also achieve ~8–10 minute gradients for sugars with good DP2–DP4 resolution. Example: VG-25 4D. 

Since the best choice depends on the sample, detection, and whether full DP resolution is critical, we encourage customers to email us with details of their experiment for tailored guidance.

Additional Comments

image_pdfDownload or Print PDF