DC613 and Na salt

Record: 25


Category: Column Usage
Product: DC-613
Sample:


Question
bought DC613 column without noticing that contains Na salt. Since using MS, want to remove Na, is it possible to replace it by ammonium ion (using ammonium formate as eluent)

 
 

Answer
First of all, I’m sorry to say but we’d like to mention that the usage of ammonium formate is not under manufacture’s guaranteed eluent for DC-613. Since it will probably change column’s separation performance, but we do not have any experience .

The application specialist mentioned that by even using ammonium formate, Na+ ion will not be replaced by NH4+ easily (i.e., probably you will not see an interference problem in MS). I am not sure what type of ionization system you are using, but can you add ACN to the mobile phase? She mentioned that you may mix a small amount of ACN to ammonium formate and use this to pre-condition the column until the baseline becomes stable. So we would recommend you to try this first.

But if you prefer to replace Na+ completely to NH4+, you may try the following.
Wash the column with acid (the working pH range for this column is between 2-14, so I would suggest to use acid pH slightly lower than 2 to remove Na+). Then rinse the column with DI water. Then use ammonium (aq) solution to add NH4+ ion. Change to the mobile phase. Again, I’m sorry but we are not certain how this process will affect the packing gels.

 
 
Additional Comments

image_pdfDownload or Print PDF