Dimethicone (~10,000 Da) – Recommended Shodex Column

Record: 729


Category: Column Selection
Product: GF-310 
Sample: dimethicone 


Question
Is the GF-510 column suitable for separation of dimethicone (~10,000 Da)?

Answer
For ~10 kDa dimethicone, the better fit is Asahipak GF-310 HQ (not GF-510). GF-310 HQ’s calibration window (~200–40,000 Da) centers on 10 kDa and typically gives sharper peaks around that mass.

If you’re running a purely organic GPC workflow (e.g., THF/toluene/chloroform), consider the KF series—purpose-built S-DVB GPC for hydrophobic polymers (silicones, resins). For a 10 kDa target KF-802 is commonly used.

When to choose which

  • GF-310 HQ → best when you want aqueous/organic flexibility (mixed-solvent SEC) and resolution right at ~10 kDa.
  • KF-802  → best for all-organic GPC of silicones; often delivers crisper peaks and very stable baselines in toluene/THF.

Reference applications: