Skip to main content
Print

Recommended Column for Dimethicone (PDMS) Separation in THF

Record: 734


Category: Column Selection
Product: KF-803L
Sample: Dimethicone


Question

A customer selected the GF-310 HQ column for Dimethicone separation because their solvent is THF. Is this the correct choice?

Answer

While the Asahipak GF-310 HQ column is chemically compatible with THF, the Shodex GPC KF-803 or KF-803L are the preferred columns for Dimethicone (PDMS) analysis.

The GF-HQ series uses a polyvinyl alcohol (PVA) matrix designed for both aqueous and mixed-organic SEC. Although THF is listed as a usable solvent, these columns are optimized for hydrophilic polymers and may show lower stability or recovery when used for hydrophobic samples such as silicone oils.

The KF-800 series, by contrast, is packed with cross-linked polystyrene–divinylbenzene (PS-DVB) gel specifically engineered for 100% organic GPC in THF, toluene, or chloroform. This makes the KF-803 an excellent fit for Dimethicone, offering better recovery, reproducibility, and column lifetime. The KF-803L provides a slightly broader linear calibration range and is also well suited for silicone polymers with wider molecular-weight distributions. Recommended application: Polystyrene Standards – Shodex GPC KF-803L

Additional Comments

image_pdfDownload or Print PDF