Question
We have GF-310 and GF-7M columns purchased about 10 years ago. Can these be used for organic acid analysis?

Answer
The GF-310 and GF-7M are aqueous GFC (size exclusion) columns for polymers and polysaccharides, not ionic compounds.
They are not suitable for organic acid analysis.

For this application, use the RSpak KC-811, which operates in ion-exclusion mode with a sulfo-modified styrene–divinylbenzene polymer base, providing excellent separation of organic acids such as lactic, acetic, citric, malic, fumaric, and succinic acids.

Example method and chromatograms:
Influence of Acidic Eluent Selection on Organic Acid Analysis (KC-811)

For LC–MS or mixed-buffer systems, RSpak DE-213(2.0 × 150 mm) can be used as a polymer-based reversed-phase alternative compatible with volatile buffers and organic solvents.