Question
We have a customer looking for two GPC columns. He would like both a water based, and organic based column. The solvents they will be using are MeOH, THF and Dichloromethane. He is looking to work with large amounts of sample, from 100-300 mg and the MW distribution is from 100-10,000. He would like to know what is available for columns in the size range of 20 mm diameter.

 
 

Answer
As you can see in our catalog pg 45-46 (http://shodex.net/data/downloads/Shodex_2011-2012_Catalog.pdf), we have two types of GFC (aqueous base) SEC columns. KW series is a silica base with polymer coating and SB series is a polymer based column. If the customer requires 100% MeOH mobile, I’d recommend KW series, if it’s <50% MeOH SB columns can be also used. Which one to choose from is depending on the analyte and other mobile phase condition (especially pH). So if you are not sure, please let me know the details of customer’s analyte and detailed mobile phase requirements. For the GPC (organic solvent) SEC columns, our KF series (pg 52) can be used with THF and dichloromethane. Although it is recommended to have one column dedicated to one solvent usage…This is also available in preparative size (pg86). The pore size that they probably need is KW/SB/KF-802.5 (2002.5 in preparative) for the analysis of compound with MW 100-10,000 (pullulan/PS). Please refer to the calibration curves on each product (shown on the right side page of the catalog). All KW,SB and KF series, the preparative size is 20x300mm. Please note that they need approximately 5 weeks to be prepared. Generally, we recommend the user to test with analytical (standard size) column first before moving on to the preparative size. Please let me know if you have further questions.     Additional Comments

image_pdfDownload or Print PDF
Shopping Cart
➤
Scroll to Top
Select more than one item for comparison.
Skip to content