GPC standard

Record: 750


Category: Specifications
Product: KF
Sample:


Question
I am using few KF series columns with THF now. Can I switch the solvent (Mobile phase) to DCM ( dichloromethane) .

 
 

Answer
Unfortunately, we do not guarantee the use of dichloromethane with the KF-800 series columns.
Since boiling point of dichloromethane is around 40 oC, if you are using a regular GPC system, you will be running the analysis very close to this temperature, which causes unstable baseline (especially with RI detector).
At our lab in Japan, they have used dichloromethane with KF-800 series, thus we know it will not damage the column, however, as I mentioned, this will be outside of our guarantee.

 
 