Question
I am wondering if you sell a Guard column in the KF series that is suitable for big molecules and especially for use with the KF-807? similar concept to the SB-807G in the aqueous series.

 
 

Answer
Compared to SB-G, the column size of KF-G is shorter (and with the smaller i.d. makes the column volume much smaller than that of SB-G). Probably because of this, they never experienced shear degradation of large compounds. Also there is hardly any back pressure on SB-G (At 1.0 mL/min flow, 0 Kg), thus there will be no back pressure issue. The pore size used is big enough and will not interfere or influence the separation of large molecules analyzed by KF-807.

Therefore, we do not think there will be any problem using KF-G with KF-807. However, if the customer really worry about it, we can recommend UT-G.

 
 