Record: 339


Category: Column Selection
Product:
Sample:


Question
I have a prospect (currently using waters Gpc).

They have these types of samples and are looking for a column to fully resolve a MW range from 0-40,000.

We will be analyzing raw materials with the GPC such as; cycloaliphatic epoxies, phenolic resins, solution vinyls, epoxidized polybutadienes, BPA epoxy, nitrocellulose, TMPTA, etc…

 
 

Answer
I would recommend KF-803 or KF-803L (KF-803 will provide slightly better separation, while KF-803L will be preferred for MW distribution analysis).
Depending on the degree of separation that the customer is looking for, they may need multiple number of KF-803 (or in combination with smaller pore-sized KF series columns).

Below are some application examples using KF series column:

Epoxy resin
http://shodex.net/index.php?seitenid=1&applic=463

Phenol resin
http://shodex.net/index.php?seitenid=1&applic=1149

Polyvinyl
http://shodex.net/index.php?seitenid=1&applic=1154

Polybutadiene
http://shodex.net/index.php?seitenid=1&applic=1152

As long as the sample dissolve in THF, the same condition can be used.
Do you know which Waters column they are currently using?

 
 