IMP – tECHNICAL QUESTION

Record: 272


Category: Column Selection
Product:
Sample:


Question

 
 

Answer
1. The question was that they are using UV detector , and they wanted to know the advantages of using our NN-614 from novapak. do you think they could use the multimode for purification of betains ?

I’ve already wrote those…

2. The customer has in stock the NN-614 and the YS-50 ( you recommended the NN 814 ). So inolab is asking me if they could supply the NN-614 instead of the NN 814 or the YS-50.what do you think ?

I thought we already discussed this too. NN-614 will probably (can never say for definite) separate. But they are using NOVAPAK, which is a preparative size column (but 8.0mm id) that’s why I thought they may want to use NN-814 instead of NN-614.

3. Since the YS-50 (125mm) is shorter than the NN-614 (150mm ) , do you think we could recommend the YS-50 ?.

YS and NN are different separation modes. Cation exchange columns are generally shorter than other types. I feel NN will have a better chance.

4. Ana spoke with the customer that they will supply them the NN-614 ( they explained that it was shorter ) and the customer said he would like to try that one regardless of the size. So remember you send them some applications with these betains ? , well this customer was asking that the applications you sent were with shorter betains but they are looking for longer betains , so do you think it will be fine to use this column NN 614 ?

The separation is based on RP and IE, longer C chain means that they will be retained longer.. That’s probably why they are seeing the long retention time.
That’s another reason C8 could be an alternative.

 
 