Question
I recently purchased a NI-424 column for non-suppressed anion chromatography and currently I am interested in sulfate analysis. I know that in suppressed systems atmospheric carbon dioxide gives rise to significant carbonate contamination, which interferes with sulfate determination. Is this a concern in non-suppressed IC?

 
 

Answer
By using NI-424, we can detect sulfate without the effect of carbon dioxide. Refer to the application:
http://www.shodex.net/index.php?seitenid=1&applic=546.

 
 