Question
We use Shodex KW-803 and KW-804 in series for PEG (5K–80K Da, NHS end group) separation. Could you explain the principle and lot variation, and suggest any replacement models?

Answer
The KW series are aqueous SEC columns packed with silica particles coated with hydrophilic polymer gel, providing high efficiency, low adsorption, and excellent reproducibility for proteins and PEGs.

Small variations between gel lots can slightly change the pore-size distribution, leading to minor shifts in elution volume—most noticeable for high-MW PEGs.

While there’s no newer version, the existing KW-803 + KW-804 combination remains standard for PEGs in the 5 K–80 K Da range.
If a silica-free alternative is preferred for higher chemical stability or alkaline buffers, consider OHpak SB-803 HQ or SB-804 HQ for polymer-based SEC approach.