I believe you are using Shodex SB-802 HQ and SB-802.5 HQ columns.
You may use nitric acid to adjust NaNO3 solution. However, NaNO3 solution will not have buffering capability.
Thus, it might be better to use NaNO3 with buffers such as phosphoric, acetic, or citric acid.
And yes, pH= 3 is the lowest working pH that we would recommend.

 
 