NH2P column technical notebook

Record: 280


Category: Column Problem
Product:
Sample:


Question
having SOS retained on column and no elution..

 
 

Answer
I received some advices from our application specialist in Japan.
We are wondering if you can use phosphoric acid (pH=2) as a mobile phase instead.
When using an amino column in an ion-exchange mode, all the amino function group needs to be ‘modified’ with counter-ions in order to stabilize the retention.
Seems like this is not done properly with current ammonium sulfate mobile.
Her advice is to equibrate the column with 60 mM phosphoric acid (pH=2), using 0.5mL/min flow rate for 2 hours and then use this solvent as a mobile (so you may increase the flow rate ~1 mL/min). If this concentration seems like not strong enough, you may use up to 0.1m phosphoric acid.
If this doesn’t work, we may need to consider using sulfuric acid as a counter ion, but sulfuric acid can react with column housing (SUS), it cannot be used at a high concentration.

 
 
Additional Comments

image_pdfDownload or Print PDF