Question
Will using a Shodex Asahipak ODP-40 (4.6 × 150 mm, 4 µm) column increase the system pressure of my HPLC?

Answer
No significant increase in pressure is expected.
The Asahipak ODP-40 is a standard HPLC column, not a UPLC column. It uses 4 µm polymer-based packing (polyvinyl alcohol with octadecyl functional groups) and operates within normal HPLC backpressure limits.

Typical specifications:

  • Max Pressure: 9 MPa (≈ 1300 psi)

  • Particle Size: 4 µm

  • Pore Size: 250 Å

  • Normal Flow Rate: 0.3 – 0.6 mL/min

  • Max Flow Rate: 0.8 mL/min

  • Column Size: 4.6 × 150 mm

  • Functional Group: Octadecyl

Because it is polymer-based, ODP-40 generally produces lower backpressure than silica-based C18 columns of similar dimensions. Pressure may vary slightly with solvent viscosity, flow rate, and temperature, but it remains well within standard HPLC operating ranges.

Additional Comments

The ODP-40 is an older model that may be discontinued in future catalog updates. For new applications, use Asahipak ODP-50 4D (4.6 × 150 mm, 5 µm), which offers similar polymeric reversed-phase selectivity, wider pH tolerance (2–12), and ready U.S. availability.
View ODP-50 4D column