Skip to main content
Print

Replacement for Agilent Asahipak ODP-50 (4.0 × 125 mm, 5 µm)

Record: 743


Category: Column Selection
Product: ODP-50 4D
Sample: N/A


Question
I have been using an Agilent Asahipak ODP-50 column (4.0 × 125 mm, 5 µm; part #79923OP-564) for several years, but it has been discontinued. I understand Shodex is the original manufacturer of the packing material. The Agilent column lists a pore size of 80 Å, while Shodex ODP-50 columns are specified as 250 Å. Is this correct, and what is the best replacement?

Answer
The Shodex Asahipak ODP-50 and Agilent-branded ODP-50 columns use the same packing material — a polymer-based reversed-phase (C18-like) polyvinyl alcohol phase. However, Shodex no longer offers the 4.0 × 125 mm format due to housing discontinuation.

Alternatively, we recommend switching to the standard ODP-50 4D(4.6 × 150 mm, 5 µm), which is readily available from U.S. stock and provides equivalent selectivity and performance.

The ODP-50 series pore size is 250 Å, so the 80 Å specification in the Agilent catalog was likely an error.

Additional Comments

image_pdfDownload or Print PDF
Shopping Cart
➤
Scroll to Top
Select more than one item for comparison.
Skip to content