Record: 192


Category: Specifications
Product: SB-803
Sample:


Question
Could I use this SB-803 HQ column in DMF system? What is the GPC column you used for the organic system, such as DMF?

 
 

Answer
Yes, SB-803 HQ can be used with 100% DMF (Just for your information, SB-802 HQ cannot be used with DMF).
Please use slower flow rate (<0.5mL/min) when you replace the eluent. We also have GPC column filled with DMF as shipping eluent. Compared to SB series, packed with polyhydroxymethacrylate gel, GPC KD series column is packed with styrene divinylbenzene copolymer. Styrene divinylbenzene copolymer is more hydrophobic than polyhydroxymethacrylate. Generally speaking, good for non-polar substance analysis. We also have GPC columns filled with THF or Chloroform. Please take a look at pg 62 on the below catalog for our column selections and applicability of their solvent usage. http://shodex.net/data/downloads/Shodex_2011-2012_Catalog.pdf     Additional Comments