Record: 274


Category: Column Selection
Product: SEC
Sample: PEG


Question
a customer wants to separate PEG’s on a 1 x 50mm column. Before I go back to him, I know of no way he can do this on so small (short) a column or by any mode other than size exclusion. Is there?

 
 

Answer
Yes, I would stay with SEC mode, depends on the MW size, but if it is smaller than 1.3MDa, I would recommend KW400 series over SB series.
Required lengths is depends on how much separation efficiency the customer requires…so I’m sorry but it is difficult to say if 50 mm is going to be long enough or not.

Below is an example on GS-320 semi-micro size column. You will see sharper peaks with narrower i.d. columns. Though there’s no change in retention time, meaning shorter the column, peaks will come out closer.
http://shodex.net/index.php?seitenid=1&applic=730
http://shodex.net/index.php?seitenid=1&applic=731

Also please note that 1.0 mm i.d., column cannot be used with a conventional HPLC system.

 
 