protein analysis Celgene Thank you for your interest in Shodex products

Record: 285


Category: Column Selection
Product: GF/GS
Sample: protein


Question

 
 

Answer
Because of the MW size of BSA, we don’t think using smaller pore sized column (i.e, 320) will separate them well either.. And I had considered the potential of combining GF/GS+KW. We cannot say for 100%, but it is less likely this will provide a desirable separation: since compounds will be separated under one separation mode, then when this mixture is going through other separation mode, this produces more peaks and may cause overlapping from the first separation mode peaks… unless this is done in two dimensions (LC/LC with column switching method). It will not be favorable, but I am thinking you may need two methods for the separation of the ionic protein and separation of BSAs.. Have you tried any other separation modes like reversed phase/ ion exchange for the ionic protein analysis?

We have technical notebook available for KW400 series. However, basic separation mode is the same as KW-800 series. So using KW400 series will create the same problem.
http://shodex.net/data/downloads/Shodex_Technical_Notebook_5_KW400.pdf

 
 