Question
A customer asked about an equivalent column for: Rezex RHM Monossacaride (300×7.8mm) da Phenomenex: to analyze insulin, glucose and fructose in food.

 
 

Answer
The equivalent will be SUGAR SH1011. (Rezex RHM and SH1011 are both under USP L17)
http://shodex.net/index.php?seitenid=1&reiter=r_results&applic=1085
http://shodex.net/index.php?seitenid=1&applic=1089
It is good for the separation of glucose and fructose, but we don’t have application for insulin..
I think insulin will be excluded first as it is much bigger than sugars.

 
 
Additional Comments
However, inulin is a polysaccharides while glucose and fructose are monosaccharides. They are not usually analyzed together. Does your customer only need to separate and quantify three compounds? Or do they need to determine the MW of inulin?
If for quantification, using SH1011, inulin will be excluded as a void and separated from other two. Although, as you found the application, for inulin analysis, NaNO3 is used as a mobile phase. I cannot say for sure that H2SO3 solvent used for SH1011 will be a best mobile for inuline.
Are they already using a Rezex column? SH1011 should give similar result in that case.