SB-806 2D Liposome analysis Thank you for your interest in our Shodex columns

Record: 31


Category: Applications
Product: SB-806 2D
Sample: liposome


Question
1. we use this column to analyze how much nonencapsulated chemicals(free drug) in our liposome sample system.
2. by using Shodex SB806-2D, we can get two peaks. one is liposomes and another one is free small molecular chemicals. the result is good.
3. we prepare sample(10ul) by using water dilute 100 times to 1ml. the anayltical conditions: mobile phase: 100% H20, temperature: 60 deg. flowrate: 0.065ml/min.
I would like to describe our problem again. now we use Waters UPLC which have PDA and ELSD detectors. after we ran our sample, we found some lipids to stick in this column. and we have to use actonitrile to clean column, it usually takes 1 hour. we tried using HBS(0.08M NaCl and 0.02M Hepes) as mobile phase, but it has problem about ELSD detector because of salt. so we prefer to use water as mobile phase. now we are looking for other columns and hope solve this problem.

 
 

Answer
I was searching publications regarding liposome analysis and found the following [C.Grabielle-Madelmont, S.Lesieur, M.Ollivon. J.Biochem.Biophys.Methods 56 (2003) 189-217].
It indicates that “presaturation”, saturate the gel with lipid (small unilamellar liposomes, SUV), provided a good recovery and integrity of the liposomes.
Are you familiar with this technique?
I think what you are seeing now is that the liposomes being adsorbed onto the surface of the gels. So once you saturate the parts that tends to retain lipids, it should be able to prevent liposome from being retained. But saturation needs to be done with small liposome, so it does not clog the column.
Please let me know how this idea sounds..

 
 